Comparison of a New Bisphenol-a-Glycidyl Dimethacrylate-Based Cortical Bone Void Filler with Polymethyl Methacrylate
The European Spine Journal (2001) Volume 10, S147-S152

Erik M Erbe, Ted D Clineff, Gina Gualtieri

Study Design. Biomechanical comparison of biocompatibility and interfacial bond strength of Cortoss and PMMA.
Abstract. A newly formulated and reinforced bisphenol-a-glycidyl dimethacrylate (bis-GMA) resin (Cortoss/Orthovita, Malvern, Pa.) was compared with Simplex P polymethyl methacrylate (Stryker Howmedica Osteonics, East Rutherford, N.J.) in rabbits for up to 52 weeks and in sheep for up to 78 weeks. As seen in scanning electron microscopy and histology examinations, both implant materials were surrounded by bone at late time periods, with fibrous layers of connective tissue seen in half the Simplex P specimens. No clinically significant safety differences between implant materials were apparent. Interfacial bond strengths between the implant and bone generally increased with time, but were 4.5-fold greater with Cortoss than Simplex P at 24 weeks. Forces required to displace 316SS rods held in place with Cortoss were consistently greater than forces to displace rods held in place with Simplex P. No statistically significant differences in displacement forces were found between rods held in place with Cortoss polymerized in situ and rods held with pre-polymerized Cortoss. Interfacial bond strengths were greater for Simplex P that was polymerized in situ than for pre-polymerized polymethyl methacrylate specimens. Cortoss synthetic cortical bone void filler is a good candidate material to fix implants in bone. It has characteristics consistent with long-term safety and has a better ability to bond to bone than Simplex P.