Mechanical Characterization of Three Percutaneous Vertebroplasty Biomaterials
The of Journal Materials Science: Materials in Medicine
(2006)
Volume 17,
421-426
Sabina
Gheduzzi, Jason
J.C.
Webb, Anthony
Miles
Study Design.
Biomechanical comparison of three vertebroplasty biomaterials.
Abstract.
Percutaneous vertebroplasty (PVP is gaining popularity for the treatment of vertebral compression fractures. The need of obtaining low viscosity materials for injection through small bore needles and the necessity of visualizing the fluid flow during injection have led users to the formulation of a number of ad-hoc recipes aimed at adapting PMMA cements for this use. Industry, on its part, has addressed these requirements by developing specific products for this application.
This study aimed at providing a direct comparison of a wide range of mechanical properties between three commercially available biomaterials developed for PVP: two PMMA based materials, Osteopal V (Merck Biomaterial GMBH, Dermstadt, D) and Vertebroplastic (DePuy Acromed, Inc, MA, USA), and a bis-GMA composite, Cortoss (Orthovita, PA, USA). Cortoss consistently exhibited higher values for compressive strength, bending modulus, and shear strength to both Osteopal V and Vertebroplastic. The creep behavior of Cortoss was also different from that of the two PMMA cements.
PVP can take advantage from the development of new injectable biomaterials in response to the problems associated with the use of PMMA in a highly vascularized area such as the vertebral body. In addition careful modulation of the mechanical properties of the material has the potential to further improve the outcome of PVP, possibly reducing the risk of adjacent level fractures associated with the procedure.